The LMP91200 device is a sensor AFE for use in low-power, analytical-sensing applications. The LMP91200 is designed for 2-electrode sensors. This device provides all of the functionality needed to detect changes based on a delta voltage at the sensor. Optimized for low-power applications, the LMP91200 works over a voltage range of 1.8 V to 5.5 V. With its ex.
•1 Active Guarding
• Key Specifications
Unless otherwise noted, typical values at TA = 25°C, VS = (VDD-GND) = 3.3 V
– pH Buffer Input Bias Current (0 < VINP < 3.3 V)
– Maximum at 25°C: ±125 fA
– Maximum at 85°C: ±445 fA
– pH Buffer Input Bias Current (
–500 mV < VINP
– VCM < 500 mV), VS = (VDD
– GND) = 0 V
– Maximum at 25°C: ±600 fA
– Maximum at 85°C: ±6.5 pA
– pH Buffer Input Offset Voltage: ±200 µV
– pH Buffer Input Offset Voltage Drift: ±2.5 μV/°C
– Supply Current: 50 μA
– Supply Voltage: 1.8 V to 5.5 V
– Operating Temperature Range:
–40°C to 125°C
– Package: 16-Pin TSSOP
2 Applications
• pH.
No. | Partie # | Fabricant | Description | Fiche Technique |
---|---|---|---|---|
1 | LMP91000 |
Texas Instruments |
programmable analog front-end | |
2 | LMP91000 |
National Semiconductor |
Sensor AFE System | |
3 | LMP91002 |
Texas Instruments |
programmable analog front-end | |
4 | LMP91050 |
Texas Instruments |
Configurable AFE | |
5 | LMP91051 |
Texas Instruments |
Configurable AFE | |
6 | LMP91300 |
Texas Instruments |
Industrial Inductive Proximity Sensor AFE | |
7 | LMP90077 |
Texas Instruments |
Sensor AFE | |
8 | LMP90077 |
National Semiconductor |
(LMP90077 - LMP90080) Sensor AFE System | |
9 | LMP90078 |
Texas Instruments |
Sensor AFE | |
10 | LMP90078 |
National Semiconductor |
(LMP90077 - LMP90080) Sensor AFE System | |
11 | LMP90079 |
Texas Instruments |
Sensor AFE | |
12 | LMP90079 |
National Semiconductor |
(LMP90077 - LMP90080) Sensor AFE System |